Back to Product Guide - Temperature Instrumentation
3 1/2 DIGIT RTD METERS & CONTROLLERS
Q2000M () / Q2000R (0.1°)

Made in the USA
Electromagnetic Compatibility
Year 2000 Compliant


Price List
Mechanical Reference
1/8 DIN (1A) Case
Mechanical Reference
Connections Diagram
QUANTA STANDARD FEATURES
100 platinum RTD input
2,3 or 4-wire connection
±1,999-count display span
1°C or 1°F resolution (Q2000M)
0.1°C or 0.1°F resolution (Q2000R)
Sensor-break detection
120 dB CMR to 500 V
1 mV/count linearized analog output
LED or LCD display
Display hold and test
120/240 Vac, 5 Vdc, 9-32 Vdc or 26-56 Vdc power
Screw-terminal barrier strip
1/8 DIN case

OPTIONS
Analog output for user-selected span
Single-setpoint 10 A relay control
Dual-setpoint 10 A relay control
Proportional 4-20 mA control
Time-proportional 2 A solid-state relay control
Isolated parallel BCD output
NEMA-4 splash-proof lens cover

QUANTA Models Q2000M and Q2000R are low-cost, 3 1/2 digit indicator/controllers for 100 ohm platinum RTD's with alpha of 0.00385. They are complete with sensor excitation and break detection. Resolution can be 1° (Q2000M) or 0.1° (Q2000R). Readout can be in °C or °F, as selected by plug-in jumpers. What makes QUANTA exceptional is the wide choice of options.

INPUT POWER AND DISPLAY OPTIONS
Six types of meter power are available: 120 Vac, 240 Vac, 24 Vac, 5 Vdc, isolated 9-32 V dc and isolated 26-56 Vdc. An LED display is standard. An LCD display is optional and is recommended for viewing in bright ambient light.

ANALOG OUTPUT OPTIONS
A 1 mV/count (±2 V full-scale) analog output is standard and is ideal for driving a strip-chart recorder. An additional analog output can be provided by an optional vertical plug-in board. Available output signals are 0-5 Vdc, 0-10 Vdc, 0-1 mA (source or sink) and 4-20 mA (source or sink). The top and bottom of each output range can be scaled to fit a user-selected display span.

CONTROL OUTPUT OPTIONS
Additional outputs can be provided by a horizontal upper board. Available options include single-setpoint control with one 10 A relay, dual-setpoint control with two 10 A relays, 4-20 mA proportional control (source or sink), time- proportional 2 A solid-state relay control, and isolated, parallel BCD output. For additional information, please refer to the QUANTA, Options, and Mechanical sections.

SPECIFICATIONS

ANALOG INPUT

Temperature RangeRTD type: 100-ohm platinum with alpha of 0.00385
Calibration: To DIN 43760, October 1980
Excitation current:
0.42 mA (Q2000M), 4.2 mA (Q2000R)
Sensor connection: 2, 3 or 4 wires
Max lead resistance: 20 ohm for 2, 3 or 4-wire connection
Zero adjustment: ±5°C (±9°F)
Overvoltage protection: 15 Vp

NOISE REJECTION
NMR: 50 dB, 50/60 Hz
CMR: 120 dB
CMV: 1500 Vp per HV test, 354 Vp per IEC spacing

ACCURACY AT 25°C
Zero tempco (bridge-network error): ±0.05 deg/deg
Span tempco, 10°C to 40°C: ±0.006% of reading/°C
Lead resistance effect
2-wire connection: +2.6°C (4.7°F)/ ohm
3- or 4-wire connection: compensated to 20 ohm max

DISPLAY
LED: 0.56 in (14.2 mm) 7-segment, red
LCD: 0.50 in (12.7 mm) 7-segment liquid crystal
Symbols: ­1.8.8.8
Decimal point positions: Three positions selectable by jumpers behind lens or at connector
Sensor-break indication: Three-least significant digits blanked

POWER
AC voltages:
120, 240 or 24 Vac +10%/­15%
AC frequency: 49-440 Hz
DC voltages: 9-32 Vdc, isolated to 300 Vp; 26-56 Vdc, isolated to 300 Vp; 5 Vdc ±5%, non-isolated
Power consumption: 5 W max

ENVIRONMENTAL
Operating temperature:
0 to 60°C
Storage temperature: ­40 to +85°C
Relative humidity: 95% at 40°C (non-condensing)

MECHANICAL
Dimensions: Newport DIN1A (1/8 DIN) case (see Mechanical section for drawings)
Weight: 17 oz (480 g)
Case material: 94V-0 UL-rated polycarbonate



Fast Quotes on Products
HOME / PRODUCT GUIDE / TECHNICAL REFERENCE
NEWPORT Electronics, Inc. (714) 540-4914 or Contact Us
Copyright 1998, Newport Electronics, Inc. All rights reserved.